TRPP 5810 Remote Pulser/Preamplifier

    TRPP 5810 Remote Pulser/Preamplifier

    TRPP5810
     

    The TRPP 5810™ unit is a high-performance remote pulser/preamplifier dedicated to TOFD inspections and compatible with the Olympus scanner line as well as PV-100 and PV-200 weld-inspection systems.

    This remote pulser/preamplifier provides an optimum signal-to-noise ratio for TOFD inspections by combining a 40 dB preamplifier with a remote high-voltage (200 V) pulse repeater in a single, small enclosure. The TRPP 5810 supports two UT channels, enabling simultaneous inspection with one or two pairs of TOFD probes.

    The TRPP 5810 can be used as a pulser and/or preamplifier.

    TRPP 5810 as a Pulser

    • Provides an additional pulse gain in order to generate a stronger signal when needed to reveal difficult-to-detect flaws.

    TRPP 5810 as a Preamplifier

    • Provides the additional gain or broadband signal-to-noise enhancement necessary for optimum signal acquisition on thick sections of material exhibiting high ultrasonic attenuation.
    • Allows driving of long cables from remotely located sensors.

    TRPP5810

    Features

    • Withstands pulsed voltages up to 300 V for pulse widths in excess of 500 ns at PRF up to 10 kHz.
    • Dimensions (W × H × L): 57 mm × 32 mm × 90 mm
    • Weight: 300 g
    • UT connectors: 8 × LEMO 00 female
    • Water/moisture resistance: NEMA 4-IP66. Rustproof.
    • Powered by an external 12 VDC source or from the instrument.
    • Power connector: Compatible with standard Olympus umbilical cables (male Fisher 103 type)
    • Red power-on LED indicator
    • Operating temperature range: -10 °C to 60 °C

    Specifications

    Pulser

    • Remote, high-voltage (200 V) pulse repeater
    • PRF of up to 10 kHz at 100 ns per channel and up to 20 kHz at 50 ns for a single channel
    • Pulser-side outputs protected against misuse and improper connection from the instrument outputs.

    Preamplifier (Receiver)

    • 40 dB preamplifier
    • Accommodates probe frequency range from 1 MHz to 15 MHz.
    • Preamplifier-side inputs and outputs protected against misuse and improper connection from the instrument outputs.

    Options

    • Power Cable (from TomoScan FOCUS LT)
      5 m power supply cable linking to TomoScan FOCUS LT™
    • Power Cable (from OmniScan)
      10 m power supply cable linking to OmniScan®
    • Transformer and Cable
      Transformer (120 VAC to 240 VAC input to 12 VDC output) with 10 m power supply cable
    No reviews
    Product added to wishlist
    Product added to compare.

    Ši svetainė naudoja slapukus

    Mes ir mūsų partneriai naudojame tokias technologijas kaip slapukai ir apdorojame asmens duomenis, pvz., IP adresą arba naršyklės informaciją, siekdami suasmeninti jums rodomą reklamą. Tai padeda mums rodyti jums aktualesnius skelbimus ir pagerina Jūsų interneto patirtį. Taip pat tai naudojame siekdami išmatuoti rezultatus arba suderinti savo svetainės turinį. Kadangi vertiname Jūsų privatumą, prašome Jūsų leidimo naudoti šias technologijas.